-Instructions for VC-1 Verifier- T&R
Product Instructions The
Verifier
|
| 1.
OPERATING INSTRUCTIONS
Connect the voltmeter to be tested to the Verifier terminals labeled "Positive" and "Negative", observe polarity. Turn the "Voltage Select" knob on the Verifier to the position marked 1.0v. Display on the voltmeter under test will read 1.000 +/-0.002 volts. Continue to turn the "Voltage Select" knob to the 1.5v and 2.0v positions. In each case the voltmeter under test will read the selected voltage within +/-0.002 volts. To test the polarity
indicator on the voltmeter undertest, reverse the positive and negative
leads connected to the voltmeter. The voltmeter should display a minus
(-) sign and the voltage selected on the "Voltage Select" knob of the Verifier.
2. BATTERY INFORMATION The Verifier operates on 6 AA batteries. To test the batteries, turn the "Voltage Select" knob to the 1.0v position and press the "Battery Test" button. The LED will light to indicate that the batteries are good. The batteries must be replaced if the LED fails to light or the voltage at the battery connector falls below 5.5vdc. Note: Although
the VC-1 Verifier is accurate to +/- 0.002 volts of the voltage selected,
most voltmeters are only accurate to +/-0.05 volts of the voltage range.
6. SHIPPING INSTRUCTIONS All instruments being returned for repair should be sent PREPAID to either address below:
Include with shipment information the nature
of the problem, purchase order, serial number |
|
T&R Product Instructions The Verifier
|