Model SC-4
Land / Sea Configurations
This manual contains operating, maintenance and storage instructions for Tinker & Rasor’s Model SC-4 AgAg/CI Reference Electrode “Land” use.
This kit includes:
1 - Ag/CI Element with end cap
1 - Ceramic Tip Assembly (Land)
1 - Tube Body with holes (Sea)
1 - Tube Body w/o holes (Land)
1 - End Plug (Storage)
1 - Brass Weight (“B”) (Sea)
1 - 4 oz. Solution (Land & Storage)
1 - Waterproof Adapter w/lead (Sea)
2 - Instruction Manuals (Land & Sea)
1 - Carrying Case w/inserts
SEE PARTS LIST DRAWING
Optional Accessories
Tinker & Rasor Models
- CPV-2 Voltmeter Analog
- CPV-4 Voltmeter Digita
REQUIREMENTS
- 1. Voltmeter – any high input resistance 20 Megaohm or greater - see “Optional Accessories”.
- 2. Test Lead – 18 Ga. stranded wire, 8 ft supplied. (longer test lead available upon request)
- 3. Electrolyte – 4 oz. “Ocean Water” supplied - Potassium Chloride 35:1 mix is okay for storage - see “Maintenance”
PREPARATION & ASSEMBLY
- 1. Clean Ag/CI element per instruction under “Maintenance” section.
- 2. Screw end cap with Ag/CI into solid tube (part #118-061) and tighten by hand.
- 3. Fill tube with “Ocean Water” (part #049-016).
- 4. Screw ceramic tip assembly into tube and tighten by hand - check “O” rings for leaks.
- 5. Attach 18 Ga. test lead to top of electrode using knurled nut (supplied).
- 6. Attach test lead to voltmeter. *
Note:
A. Ceramic tip assembly should soak in electrolyte (Ocean Water) until thoroughly saturated.
B. Ag/CI element must be cleaned, and then soaked in electrolyte at least 15 minutes prior to first use.
*Any quality high resistance meter - 20 megohm or greater such as Tinker & Rasor Model CPV-4 Digital Voltmeter.
OPERATIONS
- 1. Press the electrodes’ ceramic tip into the soil.
- 2. Submerge the tip or the entire electrode into electrolyte.
- 3. Place a sponge soaked in electrolyte under the ceramic tip for dry soil conditions.
- 4. The ceramic tip assembly must be soaked in electrolyte until thoroughly saturated.
MAINTENANCE & STORAGE
Before Use:
- 1. The Ag/CI element must be cleaned in distilled water or mild detergent and rinsed in distilled water prior to use.
- 2. Do not touch Ag/CI element after cleaning.
- 3. The Ag/CI element should soak in electrolyte for at least 15 minutes before use to stabilize readings.
After Use:
- 1. To ensure proper readings we recommend the Ag/CI element and the ceramic tip assembly be kept moist during storage. A solid body tube and protective cover are enclosed for this sorage period as well as for use in soil readings.
- 2. For long-term storage, disassemble or for use after element and ceramic tip have been allowed to dry follow instructions above “Before Use”.
STANDARD POTENTIALS to HYDROGEN
SPECIFICATIONS - LAND
Length (125.48 mm) 4.94 In.
Diameter (Max.) ( 33.27 mm) 1.31 In.
Ceramic ( 28.49 mm) 1.12 In.
Threads 1/4-20
Temperature Range (0-900 C) 0-194o F
Shipping Weight (1.9 kg) 3.5 lbs.
INTRODUCTION
This manual contains operating, maintenance and storage instructions for Tinker & Rasor’s Model SC-4 AgAg/CI Reference Electrode “Sea” use.
This kit includes
1 - Ag/CI Element w/end cap
1 - Ceramic Tip Assembly ( Land)
1 - Tube Body with holes (Sea)
1 - Tube Body w/o holes (Land & Storage)
1 - End Plug (Storage)
1 - Brass Weight (Sea)
1 - 4 oz. Solution (Land & Storage)
1 - Waterproof Adapter w/lead (Sea)
2 - Instruction Manuals (Land & Storage)
1 - Carrying Case w/inserts
SEE PARTS LIST DRAWING
Optional Accessories:
Tinker & Rasor Models
- CPV-2 Voltmeter Analog
- CPV-4 Voltmeter Digital
REQUIREMENTS
- 1. Voltmeter – any high input resistance 20 Megohm or greater - see “Optional Accessories”.
- 2. Test Lead – 18 Ga. stranded wire, 8 ft. supplied. (longer test lead available upon request)
- 3. Electrolyte 4 oz. – “Ocean Water” supplied - Potassium Chloride 35:1 mix is okay for storage - see “Maintenance”.
PREPARATION & ASSEMBLY
- 1. Clean Ag/CI element per instruction under “Maintenance” section.
- 2. Screw end cap with Ag/CI element into tube with holes (part #118-060) and tighten by hand.
- 3. Screw brass weight (“B”) into tube and tighten by hand.
- 4. Attach 18 Ga. test lead into waterproof adapter (part #009-010) - see W-7 instructions.
- 5. Remove knurled nut from top end cap and save.
- 6. Screw top of electrode into W-7 Waterproof Adapter and tighten by hand.
- 7. Attach test lead wire to voltmeter.
- 8. Lower entire electrode into electrolyte and allow to soak for 15 minutes to stabilize readings.
OPERATIONS
- 1. Submerge the tip or the entire electrode into electrolyte.
Note:
Readings should be taken as close to the structure as possible to minimize I.R. drop. Avoid surface contaminates.
MAINTENANCE & STORAGE
Before Use:
- 1. The Ag/CI element must be cleaned in distilled water or mild detergent and rinsed in distilled water prior to use.
- 2. Do not touch Ag/CI element after cleaning.
- 3. The Ag/CI element should soak in electrolyte for at least 15 minutes before use to stabilize readings.
After Use:
- 1. To ensure proper readings we recommend the Ag/CI element and the ceramic tip assembly be kept moist during storage. A solid body tube and end plug are included for this storage period as well as for use in soil readings.
- 2. For long-term storage, disassemble or for use after element and ceramic tip have been allowed to dry follow instructions above “Before Use”.
SPECIFICATIONS - SEA
Length (125.48 mm) 4.94 In.
Diameter (Max.) ( 33.27 mm) 1.31 In.
Ceramic ( 28.49 mm) 1.12 In.
Threads 1/4-20
Temperature Range
(0-900C) 0-1940 F
Shipping Weight ( 1.9 kg) 3.5 lbs.
012-155